ANB Sensors
ANB Sensors AQ2
Part no.: AQ2 5m
Calibration-free solid-state pH and temperature sensor with no glass. AQ2 calibrates itself in the water and works in fresh and sea water, in tanks, RAS and sea cages.

Part no.: DTU-A
Surface converter that gives a 4 to 20 mA or 0 to 5 V signal from an AQ2 sensor and also links to a PC over USB like a DTU.
Prices are given in the quote we send you.
The DTU-A is a converter for installation on the surface. It turns the AQ2 sensor's digital output into a 4 to 20 mA current or 0 to 5 V voltage signal, so the sensor connects to loggers, PLCs and legacy systems that need analogue inputs. It also does the job of a DTU and links to a PC over USB.
The signal spans pH 2 to 10. A test mode steps the output through 0, 25, 50, 75 and 100 percent so you can confirm the receiving system reads correctly, for example 8 mA or 1.25 V at 25 percent.
Wiring is through a 6-pin DIN connector. The current loop takes 12 to 24 V, and ANB recommends a splice kit when connecting the DTU-A to a pigtailed sensor.
Source: the manufacturer's specification.
ANB Sensors
Part no.: AQ2 5m
Calibration-free solid-state pH and temperature sensor with no glass. AQ2 calibrates itself in the water and works in fresh and sea water, in tanks, RAS and sea cages.
ANB Sensors
Part no.: AQ2 50m
AQ2 pH sensor rated to 50 m, with a 1 m cable. The same calibration-free technology as the AQ2 5m, for deeper deployments.
ANB Sensors
Part no.: AQ2 300m
AQ2 pH sensor rated to 300 m, with a 1 m cable. For deep deployments, buoys, moorings and subsea vehicles.